High precision measurement of the air permeability of porous materials.
Fulfills DIN EN ISO 9237.
Measurement principle is absolutely free of air leakage – patent DE4217682.
High dynamic ONLINE measurements inside the production line.
Static lab measurements (lab measurement unit).
Art. No. | ||
AP3N-OS | 810130 | |
Optional: | ||
Measurement range expansion to max. 1000 cfm | 810132 |
Real measurement
The product to measure is porous and therefor permeable from all sides. These properties is related to the product thickness, without countermeasures it leads to an unwanted air leakage. Different surface properties do also highly affect the measurement results as a uncontrolled air leakage.
Measurement principle AP3N
By using a pneumatic guard, the problems described above can be solved. A concentric ring chamber is used, where the same pressure as under the measurement area is created. Because there is no flow in equal pressure relations, the ring chamber takes all unwanted air leakage.
Two separate, high-dynamic regulated air pressure systems.
Strong vacuum pumps using the ejector principle.
no rotating parts.
Maintanance-free, air-beared, high-dynamic proportional valves.
Self-compensating, high precision fine pressure sensors.
High precision air flow measurement using a Laminar-Flow-Element (LFE).
Integrated temperature and humidity compensation.